
WAZN2001 Zinc Water Analyzer
Model WAZN2001
WAZN2001 online zinc water analyser using zincon spectrophotometry, with a 0 … 2 or 0 … 5 mg/L range, ±10% accuracy, automatic colour and turbidity compensation, and 4–20 mA, relay and optional RS485 outputs.
- Brand
- TGI German Technologies
- Category
- Industrial Water Analyzer
Overview
About the WAZN2001 Zinc Water Analyzer.
The WAZN2001 zinc water analyser is an online instrument dedicated to monitoring zinc in water samples. It uses spectrophotometry with zincon as the colorimetric reagent: in an alkaline state, zinc ions in the sample react with zincon to form a stable blue complex. A UV-VIS spectrophotometer measures the colour, and the intensity of the blue complex is proportional to the zinc concentration in the water. The analyser is used in sewage and wastewater treatment, industrial water production and process monitoring in industrial facilities.
Features
- Zincon spectrophotometry with a stable blue zinc complex
- Photoelectric measurement with accurate sampling and low sample injection error
- Automatic chromaticity and turbidity compensation for real water samples
- Hybrid mixing or bubble agitation for thorough mixing of sample and reagent
- Flow path designed to prevent blocking; wetted parts tested for long service life
- Reagent leak alarm and maintenance reminders
- Self-diagnosis with on-screen alarms
- One reagent set covers 250 measurements
Applications
- Sewage treatment plants
- Wastewater treatment processes
- Industrial water production
- Process monitoring in industrial facilities
LKS supplies the WAZN2001 zinc water analyser in Malaysia, with installation, calibration and after-sales support. Contact our engineers for range selection and a quotation.
Specifications
| Model | WAZN2001 |
|---|---|
| Method | Zincon spectrophotometry |
| Measuring range | 0 … 2 mg/L or 0 … 5 mg/L (customisable) |
| Accuracy | ±10% |
| Repeatability | ±5% |
| Reagent | 1 set: 250 measurements (can be customised) |
| Analogue output | 1 × 4–20 mA (expandable to 2), max. load 500 Ω |
| Analogue input | 1 × 4–20 mA (expandable to 2), 0–5 V input |
| Relay output | 4 (flexible configuration) |
| Communication | RS485 / RS232 / USB (optional) |
| Power supply | AC 230 V, 50 Hz |
| Power consumption | 200 W (not including pump) |
| Ambient temperature | 5 … 40 °C |
| Dimensions | 1400 (H) × 405 (W) × 500 (D) mm |
| Weight | 75 kg |
Lifecycle support
Installed, calibrated and maintained.
With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.
- Installation & commissioning
- Calibration & testing
- Maintenance & overhaul
- Dismantling & replacement
Related products
Industrial Water AnalyzerWACN2001 Cyanide Water AnalyzerView details
Industrial Water AnalyzerWACR2001 Hexavalent Chromium Water AnalyzerView details
Industrial Water AnalyzerTotal Chromium Water Analyzer WACR2002View details
Industrial Water AnalyzerWACU2001 Copper Water AnalyzerView details
Industrial Water AnalyzerWAFE2001 Iron Water AnalyzerView details
Industrial Water AnalyzerWAMN2001 Manganese Water AnalyzerView details
Industrial Water AnalyzerWANH3-N 2001 Ammonia Nitrogen AnalyzerView details
Industrial Water AnalyzerWANI2001 Nickel Water AnalyzerView details
Get in touch
Need a quote or technical advice?
Tell us about your application and our engineers will recommend the right instrument or system.
- 01Share your applicationProcess, medium, range and site conditions.
- 02Engineer recommendationWe match the right instrument or system to your process.
- 03Quotation and supportSupply, installation, calibration and after-sales service.