WACN2001 Cyanide Water Analyzer
Model WACN2001
WACN2001 online cyanide water analyser that measures cyanide photometrically with automatic colour and turbidity compensation, ±10% accuracy, 4–20 mA, relay and optional RS485 outputs, for sewage, wastewater and industrial process monitoring.
- Brand
- TGI German Technologies
- Category
- Industrial Water Analyzer
Overview
About the WACN2001 Cyanide Water Analyzer.
The WACN2001 cyanide water analyser is an online monitoring instrument that gives continuous, reliable cyanide readings for water and wastewater. Under weak acid conditions, cyanide in the water sample reacts with chloramine T, and the product then reacts with isonicotinic acid to form a purple-blue compound. The colour intensity of this compound is proportional to the cyanide concentration and is measured photoelectrically. Automatic chromaticity and turbidity compensation corrects for the colour and cloudiness of real samples, while bubble agitation mixes sample and reagents thoroughly. Self-diagnosis and reagent leak alarms support long-term online monitoring.
Features
- Photoelectric measurement with accurate sampling and low sample injection error
- Automatic chromaticity and turbidity compensation algorithm
- Hybrid technology or bubble agitation for thorough mixing of sample and reagents
- Specially designed flow path to prevent blocking
- Imported core components for high reliability and long service life
- Wetted parts specially tested for a long service life
- Reagent leak alarm with maintenance reminders
- Self-diagnosis with on-screen alarms
Applications
- Sewage treatment plants
- Industrial water production
- Wastewater treatment processes
- Process monitoring in industrial facilities
LKS supplies the WACN2001 cyanide water analyser in Malaysia, with installation, calibration and after-sales support. Contact our engineers for application details and a quotation.
Specifications
| Model | WACN2001 |
|---|---|
| Accuracy | ±10% |
| Repeatability | ±5% |
| Reagent | 1 reagent set for 250 measurements (can be customised) |
| Analogue output | 1 × 4–20 mA (expandable to 2), max. load 500 Ω |
| Analogue input | 1 × 4–20 mA (expandable to 2), 0–5 V input |
| Relay output | 4, freely configurable |
| Communication | RS485 / RS232 / USB (optional) |
| Power supply | 230 VAC, 50 Hz |
| Power consumption | 200 W (excluding pump) |
| Ambient temperature | 5 … 40 °C |
| Dimensions | 1400 mm (H) × 405 mm (W) × 500 mm (D) |
| Weight | 75 kg |
Lifecycle support
Installed, calibrated and maintained.
With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.
- Installation & commissioning
- Calibration & testing
- Maintenance & overhaul
- Dismantling & replacement
Related products
Industrial Water AnalyzerWACR2001 Hexavalent Chromium Water AnalyzerView details
Industrial Water AnalyzerTotal Chromium Water Analyzer WACR2002View details
Industrial Water AnalyzerWACU2001 Copper Water AnalyzerView details
Industrial Water AnalyzerWAFE2001 Iron Water AnalyzerView details
Industrial Water AnalyzerWAMN2001 Manganese Water AnalyzerView details
Industrial Water AnalyzerWANH3-N 2001 Ammonia Nitrogen AnalyzerView details
Industrial Water AnalyzerWANI2001 Nickel Water AnalyzerView details
Industrial Water AnalyzerWATN2001 Total Nitrogen AnalyzerView details
Get in touch
Need a quote or technical advice?
Tell us about your application and our engineers will recommend the right instrument or system.
- 01Share your applicationProcess, medium, range and site conditions.
- 02Engineer recommendationWe match the right instrument or system to your process.
- 03Quotation and supportSupply, installation, calibration and after-sales service.