Industrial Water Analyzer

WANH3-N 2001 Ammonia Nitrogen Analyzer

Model WANH3-N 2001

WANH3-N 2001 online ammonia nitrogen analyser using salicylic acid spectrophotometry to HJ/T 536-2009, with a 0.01 mg/l lower limit, ranges up to 100 mg/l, automatic colour and turbidity compensation and self-diagnosis for wastewater and drinking water.

Brand
TGI German Technologies

Overview

About the WANH3-N 2001 Ammonia Nitrogen Analyzer.

The WANH3-N 2001 ammonia nitrogen analyser is an online monitoring instrument for water and wastewater. It measures ammonia nitrogen by salicylic acid spectrophotometry, working to the technical requirements of HJ/T 536-2009, and combines constant optical fibre measurement with integrated digestion and colorimetric technology. Photoelectric sampling keeps injection error low, while automatic chromaticity and turbidity compensation accounts for real sample conditions. A specially designed flow path prevents blocking, and bubble agitation mixes sample and reagents thoroughly. The analyser runs hourly or interval measurements, and its self-diagnosis, alarms and quality-control functions support reliable long-term online operation.

Features

  • Salicylic acid spectrophotometry to HJ/T 536-2009
  • Low detection limit of 0.01 mg/l
  • Constant optical fibre with light intensity compensation for small long-term drift
  • Automatic chromaticity and turbidity compensation
  • Anti-blocking flow path and bubble agitation mixing
  • Reagent leak alarm and maintenance reminders
  • Self-check and self-diagnostic functions with on-screen alarms
  • Single-channel integrated valve island, easy to dismantle and clean
  • Scan-code access to running and maintenance information
  • Quality control: parallel sample test, standard sample verification, standard recovery rate detection
  • Reverse control: automatic start by flow and pH
  • Networked for real-time checking of the instrument state
  • Certified CMC and CCEP

Applications

  • Sewage treatment plants
  • Wastewater treatment processes
  • Drinking water industry
  • Industrial process water production
  • Process monitoring in industrial facilities

LKS supplies the WANH3-N 2001 ammonia nitrogen analyser in Malaysia, with installation, calibration and after-sales support. Contact our engineers for range selection and a quotation.

  • Water quality
  • Online analyzer
  • Water & wastewater
  • Optical

Specifications

ModelWANH3-N 2001
MethodSalicylic acid spectrophotometry (HJ/T 536-2009)
Measuring range0.01 ~ 2 ~ 10 ~ 100 mg/l
Quantitative lower limit0.01 mg/l
Accuracy±10% FS
Zero drift< 0.02 mg/l
Span drift< 1%
Repeatability±3%
Water sample comparison> 2 mg/l: ±10%; < 2 mg/l: < 0.2 mg/l
Measurement intervalHourly and interval measurement; can also be started via serial port
Reagent1 set for 250 measurements (can be customised)
Analogue I/O1 × 4–20 mA (expandable to 2)
Communication (optional)RS485 / RS232 / USB / Wi-Fi / Bluetooth
Ambient temperature5 … 40 °C
Power supply230 V AC, 50 Hz
Power consumption200 W (not including pump)
Dimensions1400 (H) × 405 (W) × 500 (D) mm
Weight70 kg
CertificationCMC, CCEP

Lifecycle support

Installed, calibrated and maintained.

With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.

  • Installation & commissioning
  • Calibration & testing
  • Maintenance & overhaul
  • Dismantling & replacement

Get in touch

Need a quote or technical advice?

Tell us about your application and our engineers will recommend the right instrument or system.

  1. 01Share your applicationProcess, medium, range and site conditions.
  2. 02Engineer recommendationWe match the right instrument or system to your process.
  3. 03Quotation and supportSupply, installation, calibration and after-sales service.