DSMD1900 Digital DO Sensor
Model DSMD1900
DSMD1900 digital optical dissolved oxygen sensor using the luminescence principle, with automatic temperature compensation, a 0 … 20 mg/L range, RS485 Modbus RTU output and no membrane to replace.
- Brand
- TGI German Technologies
- Category
- Dissolved Oxygen
Overview
About the DSMD1900 Digital DO Sensor.
The DSMD1900 digital DO sensor is an optical dissolved oxygen sensor that measures oxygen by the luminescence principle, with automatic real-time temperature compensation. Unlike the amperometric method, where the silver anode wears and the electrolyte weakens over time, the optical sensor has no membrane to replace and does not consume oxygen, so it needs no minimum flow or agitation. It reports dissolved oxygen in mg/L, ppm or % saturation over RS485 Modbus RTU and is used in wastewater treatment plants, water plants, surface waters and aquaculture.
Features
- Optical luminescence measurement: no membrane replacement
- Does not consume oxygen; no flow or agitation requirement
- Automatic temperature compensation with a built-in temperature sensor
- Simple installation and calibration (air calibration and zero oxygen calibration)
- Digital RS485 (Modbus RTU) output, 12 VDC / 24 VDC supply
- IP68 / NEMA 6P waterproof construction
How it works
- A blue excitation light shines on a sensor film containing a luminescent dye
- The dye emits red light; oxygen in contact with the film absorbs part of the energy, so less red light is emitted
- A photodetector compares the emitted red light with a red reference light
- More oxygen in the sample means less red light, which gives the dissolved oxygen concentration
Applications
- Wastewater treatment plants
- Water treatment plants and water stations
- Surface water monitoring
- Aquaculture
LKS supplies the DSMD1900 digital DO sensor in Malaysia, with installation, calibration and after-sales support. Contact our engineers for selection and a quotation.
Specifications
| Model | DSMD1900 |
|---|---|
| Measuring principle | Optical luminescence |
| Measuring range (DO) | 0 … 20 mg/L; 0 … 20 ppm; 0 … 200% saturation |
| Measuring range (temperature) | 0 … 50 °C |
| Accuracy | DO ±3%; temperature ±0.5 °C |
| Pressure | ≤ 0.3 MPa |
| Power supply | 12 VDC / 24 VDC |
| Communication | RS485 (Modbus RTU) |
| Calibration | Air calibration, zero oxygen calibration |
| Materials | SUS 316L, titanium coating; O-ring fluororubber; cable PVC |
| Cable length | 10 m (can be modified) |
| Process thread | G1 |
| Dimensions | 260 mm × Ø 50 mm |
| Weight | 1.85 kg |
| Storage temperature | −15 … 65 °C |
| Lifespan | 1 year |
| Protection class | IP68 / NEMA 6P |
| Wiring | White: +12–24 VDC; black: AGND; blue: RS485 A; brown: RS485 B |
Lifecycle support
Installed, calibrated and maintained.
With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.
- Installation & commissioning
- Calibration & testing
- Maintenance & overhaul
- Dismantling & replacement
Related products
Dissolved OxygenPMDO1001 Portable Dissolved Oxygen MeterView details
Dissolved OxygenWMDO1001-FM DO MeterView details
Dissolved OxygenWMDO1001-PN DO MeterView details
Dissolved OxygenWSDO1001 Pure Water DO SensorView details
Dissolved OxygenWSDO2001 High Temperature DO SensorView details
Dissolved OxygenWSDO4051 Wastewater DO SensorView details
pHWMPH1001-FM pH MeterView details
ConductivityIndustrial Conductivity SensorsView details
Get in touch
Need a quote or technical advice?
Tell us about your application and our engineers will recommend the right instrument or system.
- 01Share your applicationProcess, medium, range and site conditions.
- 02Engineer recommendationWe match the right instrument or system to your process.
- 03Quotation and supportSupply, installation, calibration and after-sales service.