Conductivity

Industrial Conductivity Sensors

Industrial conductivity sensors for continuous online measurement in pure water, power plant, sewage and seawater treatment, with cell constants from K=0.01 to K=30 covering 0–20 µS/cm up to 30–600 mS/cm, built-in temperature compensation and RS485 Modbus RTU output.

Brand
TGI German Technologies
Category
Conductivity

Overview

About the Industrial Conductivity Sensors.

These industrial conductivity sensors measure the conductivity of water continuously and online across a wide range of industries. An AC voltage is applied between the outer two electrodes, which induces an ionic current in the solution; this current is proportional to the concentration of ions in the water and is measured to give the conductivity value. The electrodes are made of conductive materials such as stainless steel, titanium alloy or platinum, and the series is available in compression-type and pipe thread-type versions with cell constants from 0.01 to 30.

Features

  • Five cell constants (K=0.01, 0.1, 1.0, 10.0 and 30.0) to match low- or high-conductivity water
  • Double-cylinder titanium alloy structure that naturally oxidises to form a chemically passive surface
  • Anti-infiltration conductive surface resistant to all liquids except hydrofluoric acid
  • K=10.0 and K=30.0 electrodes use a large-area platinum structure resistant to strong acid and alkali, with strong anti-pollution capacity
  • Built-in temperature sensor with real-time temperature compensation; compensation element (NTC 2.252K, 2K, 10K, 20K, 30K, PT100, PT1000) specified by the user
  • RS485 output with standard Modbus RTU protocol and transmission distance up to 500 m
  • Remote setting of electrode parameters and remote calibration
  • Fast response, low maintenance cost and stable long-term operation

Selecting the cell constant

  • Low-concentration solutions need closely spaced electrodes for a stronger signal, so the cell constant is reduced to 0.1 or 0.01
  • High-conductivity solutions read more accurately with a longer path length, i.e. a higher cell constant of 10 or more
  • The electrode type must be specified at purchase according to the installation method

Applications

  • Pure water and ultra-pure water plants
  • Thermal power plant water treatment
  • Sewage treatment
  • Seawater purification
  • General water treatment
  • Acid cleaning (K=30.0)

LKS supplies industrial conductivity sensors in Malaysia, with installation, calibration and after-sales support. Contact our engineers to select the right cell constant and request a quotation.

  • Conductivity
  • Water quality
  • MODBUS
  • Power plants
  • Water & wastewater

Specifications

Measuring principleContacting conductivity electrode, AC excitation
K=0.01: range / material0 … 20 µS/cm (0 … 20 MΩ) / SS316L or titanium alloy
K=0.1: range / material0 … 200 µS/cm / SS316L or titanium alloy
K=0.01 and K=0.1 connectionPipes, hoses, flanged pipes; diameter Φ6, Φ8, Φ14 mm
K=1.0: range / connection / material0 … 2000 µS/cm / ½ or ¾ NPT / SS316L, titanium alloy and platinum
K=10.0: range / connection / material0 … 20000 µS/cm / ½ or ¾ NPT / polysulfone and platinum
K=30.0: range / connection / material30 … 600 mS/cm / ¾ NPT / polysulfone
Compressive strength0.6 MPa
Working temperature0 … 60 °C
Temperature compensationNTC 2.252K, 2K, 10K, 20K, 30K, PT100, PT1000 (user specified)
OutputRS485, Modbus RTU, up to 500 m
Mounting typesCompression-type, pipe thread-type

Lifecycle support

Installed, calibrated and maintained.

With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.

  • Installation & commissioning
  • Calibration & testing
  • Maintenance & overhaul
  • Dismantling & replacement

Get in touch

Need a quote or technical advice?

Tell us about your application and our engineers will recommend the right instrument or system.

  1. 01Share your applicationProcess, medium, range and site conditions.
  2. 02Engineer recommendationWe match the right instrument or system to your process.
  3. 03Quotation and supportSupply, installation, calibration and after-sales service.