Flush Diaphragm Pressure Transmitter
Model FDPT
Flush diaphragm pressure transmitter (FDPT) for sanitary applications, using a piezoresistive silicon oil-filled sensor behind a 316L flush diaphragm. Ranges from 0–35 kPa to 0–20 MPa, ±0.5% accuracy, IP65 and Ex ia IIC T6.
- Brand
- Bont Technologies
- Category
- Pressure Instruments
Overview
About the Flush Diaphragm Pressure Transmitter FDPT.
The FDPT flush diaphragm pressure transmitter from Bont Technologies measures the gauge, absolute or sealed gauge pressure of non-corrosive gases and liquids. Its sensing element is an HT series piezoresistive, silicon oil-filled sensor with a flush diaphragm, and an integrated amplifier circuit converts the sensor signal into a standard output such as 4–20 mA. Because the diaphragm is flush with the process connection, the transmitter is mainly used where high sanitary conditions are required, including food and beverage, medical, biological and chemical applications.
Features
- Flush diaphragm in 316L with a full stainless steel construction
- Wide measuring range, from 0 … 35 kPa up to 0 … 20 MPa
- Accuracy ±0.5%, long-term stability ±0.1% FS/year (typical)
- Various output signals: 4–20 mA, 0–10/20 mA and 0/1–5 VDC
- Adjustable on site
- IP65 protection, Ex ia IIC T6 marking and CE certification
- 15–28 VDC power supply
How it works
- Sensing element: HT series piezoresistive silicon oil-filled sensor with a flush diaphragm
- Signal conditioning: an integrated amplifier circuit converts the sensor output into a standard signal
Applications
- Non-corrosive gases and liquids
- Medical treatment
- Biological products
- Chemical industries
- Food and beverage industries
- Water pipelines
Options
- Pressure type: gauge (G) or absolute (A)
- Standard ranges from 0 … 1.0 bar to 0 … 250 bar, plus −1 … 0 bar and customer-specified ranges
- Output: 4–20 mA or customer-specified
- Process connection: 1/4" PT, 1/4" NPT, 1/2" NPT or customer-specified
LKS supplies the Bont FDPT flush diaphragm pressure transmitter in Malaysia, with installation, calibration and after-sales support. Contact our engineers for range selection and a quotation.
Specifications
| Model | FDPT |
|---|---|
| Sensing element | HT series piezoresistive silicon oil-filled sensor with flush diaphragm |
| Measurement range | 0 … 35 kPa to 0 … 20 MPa |
| Standard ordering ranges | 0 … 1.0 / 1.6 / 2.5 / 4.0 / 6.0 / 10 / 16 / 25 / 40 / 60 / 100 / 160 / 250 bar; −1 … 0 bar; customer-specified |
| Pressure type | Gauge, absolute, sealed gauge |
| Proof pressure | ≤ 1.5 × rated pressure or 110 MPa, whichever is less |
| Accuracy | ±0.5% |
| Long-term stability | ±0.1% FS/year (typical), ±0.2% FS/year (max.) |
| Power supply | 15–28 VDC |
| Output signal | 4–20 mA, 0–10/20 mA, 0/1–5 VDC |
| Pressure port | M20 × 1.5, G1/2 |
| Pressure connection (ordering) | 1/4" PT, 1/4" NPT, 1/2" NPT, customer-specified |
| Housing material | 1Cr18Ni9Ti stainless steel |
| Diaphragm material | 316L |
| O-ring | Fluororubber |
| Cable | Polyethylene |
| Insulation resistance | 100 MΩ at 100 VDC |
| Protection class | IP65 |
| Explosion protection | Ex ia IIC T6 |
| Certification | CE |
Lifecycle support
Installed, calibrated and maintained.
With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.
- Installation & commissioning
- Calibration & testing
- Maintenance & overhaul
- Dismantling & replacement
Related products
Pressure InstrumentsSmart Pressure TransmitterView details
Pressure InstrumentsDigital Pressure TransmitterView details
Flow MeterElectromagnetic Flow MeterView details
Level TransmitterRadar Level TransmitterView details
Multigas AnalyzerUV DOAS Gas AnalyzerView details
Dust & Opacity AnalyzersDust & Opacity Analyzer (Single Pass)View details
pHWMPH1001-FM pH MeterView details
DataloggerPR8000 DataloggerView details
Get in touch
Need a quote or technical advice?
Tell us about your application and our engineers will recommend the right instrument or system.
- 01Share your applicationProcess, medium, range and site conditions.
- 02Engineer recommendationWe match the right instrument or system to your process.
- 03Quotation and supportSupply, installation, calibration and after-sales service.