pH

WSPH3001 Antimony pH Sensor

Model WSPH3001

WSPH3001 antimony pH sensor for water containing hydrofluoric acid, where glass electrodes are attacked: 0–14 pH, 0–60 °C, up to 6 bar, PTFE junction, no electrolyte refilling, for semiconductor and steel industry wastewater.

Brand
TGI German Technologies
Category
pH

Overview

About the WSPH3001 Antimony pH Sensor.

The WSPH3001 antimony pH sensor measures pH in water containing hydrofluoric acid and in other media that are corrosive to glass electrodes. Instead of a glass membrane it uses an antimony sensing surface, which is sturdy and corrosion resistant. The sensor detects the concentration of H+ ions in the water and converts the electrochemical signal into a voltage, which a pH meter converts into a pH reading. A solid dielectric with a large-area PTFE liquid junction resists blocking, and a long reference diffusion path extends electrode life in harsh conditions. There is no electrolyte to refill, keeping maintenance low.

Features

  • Antimony sensing element for HF-containing water and glass-corrosive media
  • Measuring range 0 … 14.00 pH
  • Solid dielectric with large-area PTFE junction, hard to block and easy to maintain
  • Long reference diffusion channel for longer electrode life in harsh environments
  • No electrolyte refilling
  • High accuracy, fast response and good repeatability
  • PPS/PC body with upper and lower ¾ NPT threads
  • Low-noise cable, 5 m standard

Applications

  • Wastewater treatment in the semiconductor industry
  • Wastewater treatment in the iron and steel industry
  • Water containing hydrofluoric acid

Usage notes

  • Not suitable where the water contains antimony or substances that react with antimony to form complex ions
  • Keep the antimony surface clean; polish it with fine sandpaper if necessary

LKS supplies the WSPH3001 antimony pH sensor in Malaysia, with installation, calibration and after-sales support. Contact our engineers for sensor selection and a quotation.

  • pH
  • Water quality
  • Water & wastewater
  • Chemical processing

Specifications

ModelWSPH3001
Sensing materialAntimony
Measuring range0 … 14.00 pH
Temperature0 … 60 °C
Pressure0 … 6 bar
Body materialPPS/PC
JunctionSolid dielectric, large-area PTFE
InstallationUpper and lower ¾ NPT
CableLow-noise cable, direct outlet; 5 m (can be modified)
Sensor length120, 150, 210, 260 or 320 mm

Lifecycle support

Installed, calibrated and maintained.

With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.

  • Installation & commissioning
  • Calibration & testing
  • Maintenance & overhaul
  • Dismantling & replacement

Get in touch

Need a quote or technical advice?

Tell us about your application and our engineers will recommend the right instrument or system.

  1. 01Share your applicationProcess, medium, range and site conditions.
  2. 02Engineer recommendationWe match the right instrument or system to your process.
  3. 03Quotation and supportSupply, installation, calibration and after-sales service.