Humidity Transmitter

Humidity Transmitter HT

Model HT-ST-TH-XXX

Humidity transmitter HT that measures relative humidity (10–100% RH) and temperature with a thin-film capacitive sensor and Pt100, giving two-wire 4–20 mA outputs, ±2% RH accuracy and an IP65 watertight enclosure in wall, duct or separated versions.

Overview

About the Humidity Transmitter HT HT-ST-TH-XXX.

The HT humidity transmitter measures relative humidity and temperature and converts both into two-wire 4–20 mA signals for a PLC, SCADA or monitoring computer. Humidity is sensed by a thin-film capacitor and temperature by an RTD Pt100 element, housed in a compact ABS watertight enclosure. It is offered as a wall mount, duct mount or separated-probe version, accepts a wide 12–40 VDC supply and can be calibrated on site at zero and span. Typical uses include multi-point monitoring of KD rooms, where each room's reading can drive its own fuel gas damper or steam control valve.

Features

  • Fast, stable and accurate measurement with excellent long-term stability
  • ±2% RH and ±0.5 °C accuracy at 25 °C
  • Watertight ABS enclosure rated IP65
  • Compact size and easy mounting
  • Wide input voltage range of 12–40 VDC
  • On-site two-point calibration (zero and span)
  • Separate two-wire 4–20 mA outputs for humidity and temperature

Models

  • TH-102W: wall mount type
  • TH-102D: duct mount type
  • TH-103S: separated type

Applications

  • KD rooms, including multi-room systems linked to PLC or SCADA
  • Restaurants and hotels
  • Parks and commercial centres
  • Marine
  • General industries

LKS supplies the HT humidity transmitter in Malaysia, with installation, calibration and after-sales support. Contact our engineers for model selection and a quotation.

  • Humidity
  • Temperature measurement
  • Remote monitoring

Specifications

ModelHT-ST-TH-XXX (TH-102W wall, TH-102D duct, TH-103S separated)
Humidity range10–100% RH (non-condensing)
Temperature range0–100 °C (−20 … 80 °C, 0 … 50 °C, −40 … 60 °C available)
Accuracy (at 25 °C)±2% RH, ±0.5 °C
Long-term stabilityBetter than 1% RH per year (typical)
Temperature compensation±0.008% RH/°C (effect at 0% RH)
Response< 15 s (90% at 25 °C in moving air at 0.5 m/s)
Humidity sensorThin-film capacitor
Temperature sensorRTD Pt100 Ω, IEC 751, DIN 43760
Humidity output4–20 mA, 0–100% RH (linear), two-wire
Temperature output4–20 mA, 0–100 °C (linear), two-wire (scalable)
Supply voltage12–40 VDC
Sensor protectionSintered filter, part no. TR-3066 (option)
HousingABS plastic watertight enclosure, IP65
ConnectionsLiquid-tight nylon cable bushing for Ø5–10 mm cables
Ambient temperature−40 … 85 °C
WeightApprox. 150 g

Lifecycle support

Installed, calibrated and maintained.

With more than two decades in valves and instrumentation across a diverse range of industries, we keep instruments accurate and valves healthy, because valve health management is vital for plant safety.

  • Installation & commissioning
  • Calibration & testing
  • Maintenance & overhaul
  • Dismantling & replacement

Get in touch

Need a quote or technical advice?

Tell us about your application and our engineers will recommend the right instrument or system.

  1. 01Share your applicationProcess, medium, range and site conditions.
  2. 02Engineer recommendationWe match the right instrument or system to your process.
  3. 03Quotation and supportSupply, installation, calibration and after-sales service.